General:
Machine type: coordinate measuring machine with CT device
Probing systems: compact X-ray sensor Modes of operation: linear path control
Meassuring software: WinWerth® Operating system: MS Windows
Measuring Ranges:
Max. part dimensions for tomography***
– For “in the image“- measurements:
from L = 48 mm (1.9“), Ø = 50 mm (2.0“) to L = 161 mm (6.3“), Ø = 177 mm (7.0“)
– With optional Raster Tomography:
from L = 343 mm (13.5“), Ø = 95 mm (3.7“) to L = 456 mm (18.0“), Ø = 289 mm (11.4“)
Max. distance X-ray source – detector:
FDD = 500 mm (19.7“)
Dimensions and Masses:
Installation area (without instrument table):
Depth: 830 mm (32.7“)
Width: 1515 mm (59.6“)
Height: 1485 mm (58.5“)
Machine weight: 1400 kg (3087 Ibs.)
Workpiece weight: 10 kg (22 Ibs.)
– For specified Error MPE: 2 kg (4.4 Ibs.)
Maximum Permissible Error MPE*
For advanced laboratory conditions
CT Sensor
for1): P: 4,5 μm
E: (4.5+L/75) μm
SD3): (3.5+L/100) μm
No air conditioning required
CT Sensor
for2) P: 7.5 μm
E: (7.5+L/50) μm
(Where L = measuring length in mm comparable to ISO 10360 and VDI/VDE 2617)
Additional Performance Data:
Resolution rotary axis: 0,36“
Resolution: 0.1 μm (0.000004“)
Positioning speed: vmax = 150 mm/s
Acceleration: amax = 300 mm/s2
X-Ray Components:
– X-Ray Source:
Transmission target: 130 kV or 160 kV
– Detector:
– Surface area: from 60×57.5 mm2 to 233×190 mm2
– Number of pixels: from (984×1024) pixel to (2800×2304) pixel
– Pixel center distance: 50–130 μm
Supply Data**:
Voltage measuring machine: 230 V (115 V) ±10%
Frequency: 50–60 Hz
Power: max. 2000 VA
Air pressure: 5.5–10 bar
Air consumption: 3000 Nl/h
Permissible Environmental
Conditions:
Environmental air: Humidity 40%–70% rel. hum., oil free
Air contamination: max. 0.05 mg/m3
Operating temperature: 10–35 °C (50–95 °F)