Millimar C 1202 | Art.-Nr. 5312025
Product type C 1202
Display TFT color display, 110 mm (4.3″), 480 x 272 pixels
Range of digital display µm ± 999 999.9
Range of digital display inch ± 39″
Range of analog display µm ± 5000 , ± 2000 , ± 1000 , ± 300 , ± 100 , ± 30 , ± 10 , ± 3
Range of analog display inches ± .19″ , ± .07″ , ± .03” , ± .01” , ± .003” , ± .001” , ± .0004″ , ± .0001”
Resolution µm 0,01 , 0,1 , 1
Graduation mm 500 , 200 , 100 , 20 , 10 , 2 , 1 , 0,2
Graduation value inches .019″ , .007″ , .002″ , .001″ , .0002” , .0001” , .00002” , .00001”
Measuring combination +A, -A, +B, -B, A+B, +A-B, -A+B, -A-B
Features 3
Test steps 1
Dynamic functions Max, Min, TIR (Max-Min), (Max+Min)/2, Average
Statistical functions Length, Angle
Configuration Keyboard
Refresh rate 40 fps
Data transmission rate 30 Hz
Error limit, digital display 0.3% (min. 0.2 μm);
0.3% (min. 0.04 μm)*
* (with N 1702 M-HR)
Error limit, analog display 0.25% of the end scale value / 0.3% of the displayed value
Data interface USB
Hardware interfaces USB , Digimatic , Control input , RS-485 , Micro SD slot for SD / SDHC cards up to 32 GB
Control inputs Control input programmable
(functions and function sequences)
Energy supply Power source , 230 V/115 V; 50/60 Hz
IP protection category IP 42
Working temperature MIN 10 °C
Working temperature MAX 35 °C
Operating temperature Min 0 °C
Operating temperature Max 40 °C
Mahr Millimar C1202 Amplifier (5312025)
$1,900.00
Mahr Millimar C 1202 Amplifier easy to read display. The C1202 Digital amplifier offers high resolution, excellent color contrast display and analog readings with selectable scales. Operators can see both the analog scale and numeric digital readings. Display range is ±5000 µm and reading resolution 0.1 µm. The display can be adjusted to the ideal viewing angle (0º to 90º) and the C 1202 unit can also be wall-mounted. Features include: Dynamic Measurement Capability, Display to Tolerance Viewing, and User Password Setup Menu Lockout. See full details under the Info & Guides Tab.
see resource tab for required accessories for an air gage set up.
Read More
Mahr Millimar C 1202 5312025 Features
- Compact amplifier Millimar C 1202
Features: 3
Connect different sensors by way of N1700 measuring modules
Range of analog display mm: ± 5, ± 2, ± 1, ± 0,3, ± 0,1, ± 0,03, ± 0,01, ± 0,003
Range of digital display mm: ± 999,9999
Data interface: USB
Energy supply: Power source, 230 V/115 V; 50/60 Hz
Package contains: power source, micro SD memory card, instruction manual, without N 1700 module (necessary optional accessory) - Versatility for a wide range of measuring tasks
- Interchangeable N 1700 modules as measuring channels for inductive and
incremental probes or for pneumatic measuring equipment. - Perfect display thanks to high resolution and high contrast color screen
- Fully adjustable display for ideal viewing angle
- Variable display of measuring values, either as a number, pointer or bars
- 1 - 3 features can be displayed at once
- Extremely easy to operate
- Programmable switching input to control the measuring routine
- Flexible, multiple measuring tasks can be stored on the memory card.
- Compact housing
- Suitable for wall mounting
| Label | part. no. | Product type | Resolution µm | Measuring range, inductive probe | Probe inputs | Compatibility | Configuration | Data transmission rate | Error limit | Data interface | Energy supply | IP protection category | Working temperature MIN | Working temperature MAX | Operating temperature Min | Operating temperature Max | Temperature coefficient | Product weight | Zero setter | For measuring instrument | Hardware interfaces | Cable length | |
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
![]() |
Module for inductive probes N 1702 M |
5331120 | N 1702 M | 0,1 | ± 5000 , ± 2000 , ± 1000 , ± 500 | 2 | Mahr , Mahr 1340 , Mahr half-bridge , Mahr LVDT , Mahr VLDT | Millimar Cockpit Software | 4189 Hz | 0,3 % (min. 0,2 μm) | RS-485 | + 5V from N 1700 Bus | IP 42 | 10 °C | 35 °C | 0 °C | 40 °C | 0.02 µm/°C | 0.27 kg | ||||
![]() |
Module for inductive probes N 1702 M-HR |
5331125 | N 1702 M-HR | 0,01 | ± 200 | 2 | Mahr , Mahr 1340 , Mahr half-bridge , Mahr LVDT , Mahr VLDT | Millimar Cockpit Software | 4189 Hz | 0,3 % (min. 0,04 μm) | RS-485 | + 5V from N 1700 Bus | IP 42 | 10 °C | 35 °C | 0 °C | 40 °C | 0.02 µm/°C | 0.27 kg | ||||
![]() |
Module for inductive probes N 1702 T |
5331121 | N 1702 T | 0,1 | ± 5000 , ± 2000 , ± 1000 , ± 500 | 2 | Tesa | Millimar Cockpit Software | 4189 Hz | 0,3 % (min. 0,2 μm) | RS-485 | + 5V from N 1700 Bus | IP 42 | 10 °C | 35 °C | 0 °C | 40 °C | 0.02 µm/°C | 0.27 kg | ||||
![]() |
Module for inductive probes N 1702 U |
5331122 | N 1702 U | 0,1 | ± 5000 , ± 2000 , ± 1000 , ± 500 | 2 | Marposs | Millimar Cockpit Software | 4189 Hz | 0,3 % (min. 0,2 μm) | RS-485 | + 5V from N 1700 Bus | IP 42 | 10 °C | 35 °C | 0 °C | 40 °C | 0.02 µm/°C | 0.27 kg | ||||
![]() |
Module for pneumatic gages N 1701 PF-2500/5000 |
5331155 | N 1701 PF-2500/5000 | 0,1 | Mahr-Federal | Millimar Cockpit software | 4189 Hz | ± 0.8 µm in the range ± 38 µm and ± 0.5 µm in the range ± 25 µm | RS-485 | + 5V from N 1700 Bus | 0 °C | 40 °C | Electrical | ||||||||||
![]() |
Module for pneumatic gages N 1701 PF-10000 |
5331157 | N 1701 PF-10000 | 0,1 | Mahr-Federal | Millimar Cockpit software | 4189 Hz | ± 0.8 µm in the range ± 38 µm and ± 0.5 µm in the range ± 25 µm | RS-485 | 0 °C | 40 °C | Electrical | |||||||||||
![]() |
Data cable USB bi-directional DK-U1 |
4102603 | DK-U1 | USB | Micromar 40 EWR (≥ 2018) Millimar C 1202 MarSurf M310 |
USB , Duplex | 2 m | ||||||||||||||||
![]() |
Interface Adapter with Data Cable Digimatic DK-D1 |
4102606 | DK-D1 | Digimatic | Micromar 40 EWR (≥ 2018) Millimar C 1202 |
2 m | |||||||||||||||||
![]() |
Foot switch to trigger data transmission 16 ESf |
4102058 | 16 ESf |
DATA COLLECTION OF MAHR C 1202
Option 1 (Recommended): USB Keyboard Wedge
This is the method most shops use because it requires no programming.
What you need:
• Mahr DK-U1 USB bidirectional cable (Part No. 4102603).
Setup
1. Connect the DK-U1 cable from the C1202 to your PC.
2. Install the Mahr USB driver if prompted.
3. Open Excel.
4. Click in the first cell where you want data (for example, A2).
5. On the C1202, configure the USB output to transmit data.
6. Press the DATA key (or use an optional footswitch).
7. The measurement is “typed” into the active Excel cell, and Excel advances to the next row if a carriage return is enabled
Option 2: Mahr Millimar Cockpit Software
If you’re collecting inspection data regularly, Mahr’s Millimar Cockpit software is a better choice.
It allows you to:
• Display live readings
• Record measurements automatically
• Store inspection data
• Export directly to Excel (.xlsx or .csv)
• Generate reports
This is the preferred solution if you’re doing SPC or production inspections.
Option 3: Digimatic Output
The C1202 also supports Digimatic output. With a Digimatic-to-USB interface, data can be sent into Excel using software that supports Digimatic input.
Automatic Data Collection
If you’re using a footswitch (Mahr Part No. 4102058):
1. Place the part in the air gage.
2. Wait for the reading to stabilize.
3. Press the footswitch.
4. The C1202 sends the measurement to Excel.
5. Excel advances to the next row.